If the plasma does not ignite, proceed as follows:
1. Check the argon flow. Is it sufficient? Gas flow rate at least
1.5 l / min.
2. Check the connection to the HF generator. Is this OK?
3. If the argon flow and connection are OK, move the APC ap-
plicator as close as possible to the metallic object. Activate
the unit several times if necessary, until the plasma ignites.
Once the plasma has ignited, it normally ignites without delay
for subsequent activations.

6 How to use

6.1 Only use cleaned, disinfected and sterilized
products
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.

6.2 Check the product

Before each use, check the insulation, especially the ceramic in-
sulation at the distal end of the APC-applicator, for damage.
If damaged, do not use this product!

6.3 Assembling the product

Screw the APC-applicator - depending on the unit being used -
onto one of the following ERBE APC-handles:
ERBE APC
BEAMER 1
handpiece
BEAMER 2
20132-030
X
APC 300
APC 2
28