Herunterladen Inhalt Inhalt Diese Seite drucken

How To Use; Only Use Sterile Products; Check The Product; Assembling The Product - ERBE VIO ReMode Bedienungsanleitung

Elektrodengriff
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
OBJ_DOKU-184600-004.fm Seite 19 Montag, 14. September 2015 11:18 11
4

How to use

4.1

Only use sterile products

Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
4.2

Check the product

Before each use, check the product including all the insulation
for signs of physical damage!
If damaged, do not use this product!
4.3

Assembling the product

Insert the HF electrode firmly into the handle as far as it goes!
4.4

Conducting a performance test

Before every use conduct a performance test:
1. Switch on the VIO electrosurgical unit and select the basic
program.