6.4 Conduct ignition test
Check whether the plasma ignites upon first activation (see
chapter "Ignition test").
6.5 Endoscope selection
Select an endoscope for instruments with diameter 5 mm.
6.6 Using the product
Protect the APC-applicator against physical damage. Do not
bend! Do not throw! Do not use force!
7 What to do if the plasma no longer
ignites during use
1. Check the argon flow. Is it sufficient? Gas flow rate at least
1.5 l / min.
2. Check the connection to the HF generator. Is this OK?
3. Check the APC applicator. Maybe the tip has burned down.
If so, use a new APC applicator.