8.6 Required Aids

Reprocessing
Items Needed
step

Precleaning

Soft plastic brush/soft cloth
10 ml disposable syringe with male luer-
lock connection
ERBE irrigation adapter 20132-087
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
10 ml disposable syringe with male
Luer-Lock connection (4 ea.)
ERBE irrigation adapter 20132-087
Compressed air gun (2 bar maximum)
Machine cleaning/
ERBE irrigation adapter 20132-087
disinfection
8.7 Precleaning
Use water for precleaning, a non-fixing disinfectant if neces-
sary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Screw the irrigation adapter (Art. No. 20132-087) onto the
APC applicator.
3. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).