3. Start a tested program with the following properties:
Thermal disinfection: 5 to 10 minutes at 90 to 93 °C
(with a tolerance in accordance with DIN EN
ISO 17665), A
≥ 3000.
0
Final rinse with distilled or deionized water.
Sufficient product drying.
4. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning
and disinfection.
5. Remove the rinsing adapter when the program has ended.

8.10 Check

1. Check the product for visible damage and wear:
Damage to the product, e.g. cracks, rough surface,
splintering, discoloration.
Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
Damage to the coating.
Bending of the product or parts of the product. Do not
attempt to straighten bent instruments!
2. Check the mechanical functionality of the product:
Can the electrode be extended and retracted?
If damaged, do not use this product!

8.11 Packaging

1. Pack the product in disposable sterilization packaging (sin-
gle or double packaging) made from paper/polyethylene
and/or in a sterilization container.