8 Cleaning, disinfection, sterilization

8.1 Safety instructions

ERBE
Elektromedizin
disinfection. Because it is much less effective, manual cleaning
is not recommended.
Suitable for cleaning in the ultrasonic bath. Observe the follow-
ing restrictions:
Suitable for a maximum of 20 ultrasound cycles: 20132-
133, 20132-211
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.

8.2 Reconditioning limitation

Frequent reconditioning can damage this product. Observe the
safety instructions for checking the product. With visible dam-
age or functional impairment, the product may no longer be
used.