8.12 Sterilization

Only sterilize products that have been cleaned and disinfected.
ERBE Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are
used, ERBE Elektromedizin shall bear no responsibility.
Steam sterilization
Fractionated vacuum method with adequate product dry-
ing
Holding time 3 to 18 minutes at 132 to 135 °C (with a tol-
erance in accordance with DIN EN ISO 17665)
Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer
manufacturer concerning loading, handling, and drying times.