Using the RS 485 interface, the message structure remains identical to the one
NOTE
used for the RS 232 interface, the only difference being that the value assigned
to the ADDRESS <ADDR>
The controller can answers with the following response types:
Tab. 14
Response Type
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown Window
Data Type Error
Out of Range
Win Disabled
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
Response
Response
Length
Value
1 byte
-
6 bytes
-
10 bytes
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
(0x34)
1 byte
(0x35)

Technical Information

Letter Protocol Description
Description
after a read instruction of a logic
window
after a read instruction of a
numeric window
after a read instruction of an
alphanumeric window
the command execution has been
successfully completed
the command execution has been
failed
the specified window in the
command is not a valid window
the data type specified in the
command (Logic, Numeric or
Alphanumeric) is not accorded with
the specified Window
the value expressed during a write
command is out of the range value
of the specified window
the specified window is Read Only
or temporarily disabled (for
example you can't write the Soft
Start when the Pump is running)