Error Messages
Indicates a fault in the circuit of the controller used to power the
pump.
If the pump is stopped by an emergency stop signal provided via a remote
NOTE
contact, the display will be as shown in the following figure.
Remove the controller power cable and check the emergency
condition.
Then reconnect the power cable and press the START pushbutton to
start the pump.
Controller power supply circuitry is faulty, or the Controller received
a spike.
Press the START push-button twice to start the pump. Should the
message still be present, call the Agilent service.
Check connections and shortages between pump and controller, then
press the STOP RESET pushbutton twice to start the pump.