Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies Turbo-V 301-AG Bedienungshandbuch Seite 234

Rack controller
Vorschau ausblenden Andere Handbücher für Turbo-V 301-AG:
Inhaltsverzeichnis

Werbung

17
Technical Information
Letter Protocol Description
When the Request is "J" (Numerical readings), the Answer will
contain a string of 5 characters with the following parameters:
Tab. 11
BYTES
1
2
3
4
5
When the Request is "K" (Counters readings), the Answer will
contain a string of 11 characters with the following parameters:
Tab. 12
BYTES
1-4
5-8
9-10
11
On request a sample program in QBasic language is available by
Agilent.
234/294
MEANING
Current (0-255 scaled)
Voltage (0-255 scaled)
Rotational speed KRPM
Pump temperature °C (0-254 temperature reading, 255 = fail)
CRC
MEANING
Cycle time
Pump life
Cycle number
CRC
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
(F)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

969-8991969-8992969-8993Turbo-v 301-ag rack

Inhaltsverzeichnis