11.4 OPERATION FAULTS

Fault detected: The central unit does not switch on.
Possible cause
Power cord not connected
properly
Unit has not been switched
on.
Wrong voltage.
Broken connections inside.
Transformer damaged.
PCB faulty.
Filter supply defective.
Power interrupted.
Fault detected: The central unit turns off and stops working suddenly
Possible cause
Power cable damaged.
Filter damaged.
Transformer damaged.
PCB damaged.
Connections inside
interrupted.
Maximum torque is
reached.
Fault detected: The central unit doesn't switch off
Possible cause
Power filter damaged.
Fault detected: The footswitch doesn't start the motor
Possible cause
Micromotor cable not
connected properly
Footswitch cable is not
connected properly
Male connector of the
Micromotor damaged.
Pins of the connector
damaged
Wrong voltage.
Footswitch damaged.
Fault detected: It is not possible to change "Program", FWD/REV and Flow by using the footswitch.
Possible cause
Footswitch damaged.
Metal pins damaged in the
connector
Solution
Make sure the power cord is properly plugged into the electrical outlet and the connector on
the back of the unit.
Check the back of the unit to see that the On button has been pressed.
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another working electrical outlet. If the unit does
not turn on, verify the integrity of the fuses in the unit.
Immediately discontinue use of the device and contact
Solution
After checking the correct connection to the electrical outlet and the absence of liquid in the
unit, try to power the central unit. If the device does not turn on, immediately discontinue use
of the device and contact
Check if a warning message "Torque overload" is displayed on the screen. Realease the
footswitch and reactivate the micromotor.
Solution
Unplug the power cord from the electrical outlet and the central unit. Then contact
after-sales service.
Solution
Check to see that the Micromotor cable is securely plugged into the connector on the front
right of the central unit, disconnect and reconnect.
Check that the footswitch cord is plugged into the connector on the front left of the central
unit, disconnect and reconnect.
Unplug the Micromotor and make sure that the male connector is not damaged, wet or damp.
Unplug the Micromotor and the footswitch cable, make sure the metal pins on the male
connector are intact and not damaged and have not broken off and/or remained stuck in the
female connector mounted on the unit.
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another electrical outlet.
Check the ability to start the Micromotor from the central unit keyboard, pressing the [Motor
ON/OFF switch]. If so, contact
Solution
Check the possibility of changing the parameters by using the unit keyboard. Contact
®
SATELEC
after-sales service.
®
SATELEC
after-sales service.
®
SATELEC
after-sales service.
®
SATELEC
after-sales service.
®
SATELEC
19