XII - MONITORING /
MAINTENANCE
Important:
In the case of a fault, you are advised to
contact the supplier of your device. Do not use
unauthorized repairers, who might make your
device dangerous for you and your patients.

12. 1 MONITORING

Regular monitoring of the device and its
accessories is necessary in order to detect any
insulation fault or damage. Replace them if
necessary.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
12. 2 PREVENTIVE AND CORRECTIVE
MAINTENANCE
P
2 does not need any preventive
IEZOTOME
maintenance other than monitoring of the
accessories and routine cleaning/disinfection.

12. 3 FUSE REPLACEMENT

P
2 is protected by two fuses located in
IEZOTOME
the power receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device OFF (position O).
- Unplug the power cord from the electricity
supply.
- Unplug the power cord from the receptacle
(Fig. 2-1). Insert the tip of a flat screwdriver
into the notch above the fuse drawer to release
it.
- Remove the old fuses.
- Replace the old fuses with fuses of the same
type and rating.
- Push the fuse drawer back into its housing until
it clicks to indicate the correct position.
- Plug the power cord into the receptacle
(Fig. 2-1). Plug the power cord into the
electrical supply.
Note:
The device also has an internal fuse not
accessible to the user.
Contact SATELEC after-sales service (see
Chapter 2 - Repairs).
All the information necessary for the repair of
faulty components is made available by SATELEC
to the technical personnel of the SATELEC-
approved dealer network who carry out the work.
12. 4 OPERATION FAULTS
Refer to the table below: