Fault
Device does not operate
(LCD screen off)
Device does not operate
(LCD screen on)
No spray
Inadequate spray
Power fault
weak tip vibration
No ultrasonic function
No light
Leak between the
N
handpiece and
EWTRON
the handpiece cord
Liquid leak in the
irrigation pump
20
Possible cause
Power cord not plugged in properly.
Power switch in position O (OFF).
No power voltage.
Power receptacle fuse(s) blown.
Internal fuse blown
Transmission fault.
Irrigation solution bag or bottle empty.
Irrigation off.
Irrigation line clogged.
Incorrect irrigation flow rate adjustment. Adjust the irrigation flow rate.
Tip worn or distorted.
Incorrect power adjustment.
Incorrect use: incorrect angle of attack or
inadequate pressure on the tooth.
Liquid or moisture present between
handpiece and cord.
Incorrect tip tightening.
Faulty connector contact.
Handpiece cord wire cut.
Handpiece light ring absent.
Faulty light ring.
Faulty light ring connector contacts.
Light ring the wrong way around.
Faulty handpiece and/or cord connector
contacts.
Other.
Worn handpiece 1.15x1 seal
Broken tube in the irrigation line
cassette.
Remedy
Check the electric socket.
Return to SATELEC after-sales.
Place the power switch in position I (ON).
Call an electrician.
Replace the power receptacle fuses.
Return to SATELEC after-sales.
Switch the device OFF then ON again.
Return to SATELEC after-sales.
Replace the bag or bottle of irrigation
solution.
Press the irrigation ON/OFF button.
Change irrigation line.
Replace the tip.
Refer to the T
or clinical booklet for
IPBOOK
the necessary information.
Refer to the T
or clinical booklet for
IPBOOK
the necessary information.
Dry the electrical contacts thoroughly.
Retighten the tip with the wrench.
Clean the connector contacts.
Return to SATELEC after-sales to replace
the cord.
Install the light ring.
Replace the light ring.
Clean the light ring connector contacts.
Position the light ring according to the
foolproofing stud.
Clean the handpiece and/or cord connector
contacts.
Contact your installer-fitter.
Replace the seal (kit F12304).
Replace the irrigation line.