tip that has been damaged by wear or by
accidental impact (dropped, bent, etc.).
Use the T
to check scaler tip wear. Replace
IPCARD
routinely-used tips at least once a year.
b) Pre-disinfection / chemical cleaning
To be carried out immediately after treatment,
wearing thick gloves. Immerse in an ultrasonic
bath (enzymatic, quaternary ammonium base
detergent / disinfectant solution), carefully
respecting the concentration and contact time
recommended by the manufacturer of the
solution.
Use a solution that bears the EC marking or that
is in compliance with any standard required by
the national regulations.
Rinse under running water for at least 30
seconds.
c) Mechanical / chemical cleaning
To be carried out immediately after pre-
disinfection/chemical cleaning, wearing thick
gloves. Scrub the products in a new bath of
enzymatic,
quaternary
detergent / disinfectant solution, for at least 30
seconds, and in any case until all remaining
traces of contamination have been removed.
Rinse again under running water for at least 30
seconds.
d) Drying
Dry with a clean, non-woven single-use cloth, so
as to remove any traces of liquid. Pack in single-
use sterilization pouches or bags in compliance
with the specifications defined in standard EN ISO
11607-1 or in any equivalent standard required by
the national regulations.
18
ammonium
base
e) Sterilization
The tips must be sterilized individually in an
autoclave,
according
parameters:
- autoclave: type B, compliant with standard EN
13060
- sterilization temperature: 134°C
- sterilization stabilization time: 18 minutes
- pressure: 2 Bars min.
f) Storage
Store the sterilized products in a dry place
protected from dust. Before re-use, if the
integrity of the packaging is not certain,
repackage then re-sterilize in accordance with
the defined protocol.
If contamination is visible inside the bag, destroy
the product.
g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.
to
the
following