Herunterladen Inhalt Inhalt Diese Seite drucken

Piezotome Function; Newtron Function - Acteon satelec Piezotome 2 Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 70
P
function
IEZOTOME
Operation:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics:
No-load voltage: 250 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 120 ml/min. (nominal
value)
Adjustment in 10 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
N
function
EWTRON
Operation:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics:
No-load voltage: 150 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 40 ml/min. (nominal
value)
Adjustment in 1 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
Protection
Electrical equipment class: Class 1
Electrical safety class:
Type BF in P
IEZOTOME
Type BF in N
EWTRON
Safety devices
Safety shut off if internal malfunction
2 fuses (power receptacle): 5 mm x 20 mm / 2 AT
for 100 VAC to 230 VAC
1 internal fuse not accessible to the user
reference F1: 5 mm x 20 mm – 10 AT / 250 VAC
Overall dimensions
Control unit:
Width: 472.9 mm
Height: 149.5 mm
Height with bracket: 471.1 mm
mode
mode
Depth: 339.9 mm
Weight: 5 kg without accessories
Footswitch:
Width: 311 mm
Depth: 209 mm
Height: 181 mm
Weight: approximately 3.5 kg
LCD screen:
Height: 86 mm
Width: 115 mm
Scaler handpiece cord: 2 000 mm (2 900 mm
optional)
P
2 handpiece cord: 2 000 mm (2 900 mm
IEZOTOME
optional)
Temperatures
Operation: +10°C to +40°C
Storage: -20°C to +70°C
Humidity
Operation: 30% to 75%
Storage: 10% to 100% including condensation
Atmospheric pressure
Between 500 hPa and 1060 hPa
9