11. 2 DEVICE MAINTENANCE

Important:
Never use an abrasive cleaning agent on the
device.
Do not use sprays or liquids to clean and
disinfect the P
IEZOTOME
The device, the brackets and the multifunction
footswitch are not sterilizable.
The P
2 control unit, the control
IEZOTOME
footswitch and the bracket must always be
cleaned and disinfected after each procedure,
using alcohol, disinfectant agents or disinfectant
wipes for dental practices, such as S
Regular monitoring of the P
necessary in order to detect any problem.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.

11. 3 SCALER CORD MAINTENANCE

Important:
Never use an abrasive cleaning agent on the
cords.
The cords must not be immersed or disinfected
using agents containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected
and sterilized after each procedure, using
alcohol, disinfectant agents or disinfectant wipes
for dental practices, such as S
Disconnect the N
EWTRON
before cleaning, disinfection and sterilization.
2 control unit.
™ wipes.
EPTOL
2 unit is
IEZOTOME
™ wipes.
EPTOL
handpiece from its cord
b) Sterilization
The cords can be sterilized in an autoclave under
the following conditions:
- Autoclave: Class B
- Sterilization temperature: 134°C at 2 Bar
- Time at sterilization temperature: 18 minutes
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.

11. 4 ULTRASONIC HANDPIECE MAINTENANCE

After each use, the handpiece irrigation circuit
must be rinsed with distilled or demineralized
water for 20 to 30 seconds.
Cleaning and disinfection:
The handpieces must always be cleaned,
disinfected and sterilized after each procedure,
using alcohol, disinfectant agents or disinfectant
wipes for dental practices, such as S
For handpiece sterilization, refer to the manual
for each product.

11. 5 TIP MAINTENANCE

a) Lifetime
Tip shape and weight are the determining
characteristics for obtaining maximal efficiency
of the ultrasonic generator.
Careful monitoring of these two characteristics
by the user will ensure that optimal performance
of the device is maintained.
Consequently, you are strongly advised not to
alter the structure of the tips by filing or twisting
them or modifying them in any other way.
Similarly, through normal wear, tip aging leads to
alteration of its characteristics. Always replace a
™ wipes.
EPTOL
17