Herunterladen Inhalt Inhalt Diese Seite drucken

Scpi Standard Commands - orbit controls OCM160 Betriebsanleitung

Inhaltsverzeichnis

Werbung

OCM160 Precision DC Calibrator
the bits in the Event Status Register. The Event Status Enable Register is read with the common query
*ESE?. Data is returned as a binary-weighted value. The Event Status Enable Register is written to by
the common command, *ESE. Sending the *ESE common command followed by a zero clears the
ESE. The Event Status Enable Register is cleared upon power-on.
It suppresses or allows bits in ESR register. Value „0" of a bit of ESE register suppresses influence of
appropriate bit of ESR register on value of sum bit of ESB status register. Setting of any unmask bit
of ESR register results in setting of ESB status register. ESE register value can be modified by
command *ESE followed by value of mask register (integer in range 0 –255). Reading of the register
can be performed with command *ESE?. The register is automatically reset after switching on. The
register is not reset with *CLS command.
Operation Status Register
Not used in this device.
Questionable Status Register
Not used in this device.
Output Queue
The Output Queue stores response messages until they are read from control unit. If there is at
minimum one sign in the output queue, MAV register (message available) is set. The Output Queue is
cleared upon power-on and after reading all signs from output queue.
Error Queue
The Error Queue stores error messages. They are placed in a "first in, first out" queue.
The queue is read destructively using the query command "SYSTem:ERRor?" to obtain a code
number and error message. The query "SYSTem:ERRor?" can be used to read errors in the queue
until it is empty, when the message "0, No Error" will be returned.
7.8.

SCPI Standard Commands

This chapter describes standard SCPI commands.
*IDN?
Syntax:
*IDN?
Description:
This command returns the identification of the manufacturer, model, serial number and
firmware revision.
Parameters:
<CPD>
<CPD>
<DNPD>
<DNPD>
Remarks:
Overlapped command
Example:
*IDN? Response: MEATEST,OCM160,712341,1.000
*OPC
Syntax:
*OPC
41