orbit controls OCM160 Betriebsanleitung Seite 50

Inhaltsverzeichnis

Werbung

Orbit Controls AG
Example:
DISP:BRIG 1.0
DISP:BRIG? Response: 1.000000E+00
:DISPlay:LANGuage
Syntax:
:DISPlay:LANGuage <CPD>
:DISPlay:LANGuage?
Description:
This command selects language that is used on device display.
Parameters:
<CPD>
·ENGLish
Remarks:
Overlapped command
Example:
DISP:LANG ENGL
DISP:LANG? Response: ENGL
:OUTPut[:STATe]
Syntax:
:OUTPut[:STATe] <BOOL>
:OUTPut[:STATe]?
Description:
This command switches ON/OFF output terminals.
Parameters:
<BOOL>
·ON
·OFF
·1
·0
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
OUTP ON
OUTP? Response: 1
:OUTPut:LOW
Syntax:
:OUTPut:LOW <CPD>
:OUTPut:LOW?
Description:
This command connects or disconnects the Lo terminals of voltage and current outputs
to/from GND terminal.
Parameters:
<CPD>
·FLOat
·GROund
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Owner's Manual
{ENGLish} (default ENGL)
english version
{ON|OFF|1|0} (default 0)
output terminals are switched on
output terminals are switched off
same as ON
same as OFF
{FLOat|GROund} (default GRO)
OCM160 Precision DC Calibrator
50