OCM160 Precision DC Calibrator
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
TCPL:RJAM 23.0
TCPL:RJAM? Response: 2.300000E+01
[:SOURce]:TCPL:RJEXtern:TYPE
Syntax:
[:SOURce]:TCPL:RJEXtern:TYPE <CPD>
[:SOURce]:TCPL:RJEXtern:TYPE?
Description:
This command selects reference junction type. Node SOUR:TCPL also selects TC
function if not already selected.
Parameters:
<CPD>
·PLATinum
·NICKel
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
TCPL:RJEX:TYPE PLAT
TCPL:RJEX:TYPE? Response: PLAT
[:SOURce]:TCPL:RJEXtern:ZRESistance
Syntax:
[:SOURce]:TCPL:RJEXtern:ZRESistance <DNPD>[<UNIT>]
[:SOURce]:TCPL:RJEXtern:ZRESistance?
Description:
This command sets resistance at 0 °C for Platinum reference junction. Node
SOUR:TCPL also selects TC function if not already selected.
Parameters:
<DNPD>
<UNIT>
·OHM
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
TCPL:RJEX:ZRES 100.0
TCPL:RJEX:ZRES? Response: 1.000000E+02 OHM
[:SOURce]:TCPL:RJEXtern:PTSTandard
Syntax:
[:SOURce]:TCPL:RJEXtern:PTSTandard <CPD>
[:SOURce]:TCPL:RJEXtern:PTSTandard?
Description:
This command selects Platinum temperature standard for reference junction. Node
SOUR:TCPL also selects TC function if not already selected.
59
{PLATinum|NICKel} (default PLAT)
Range 100.0 ... 200.0 (default 100.0)
{OHM}