Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
VDC:LIM:RES ADJ
VDC:LIM:RES? Response: ADJ
[:SOURce]:CDC[:AMPLitude]
Syntax:
[:SOURce]:CDC[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:CDC[:AMPLitude]?
Description:
This command sets current amplitude in Current function. Node SOUR:CDC also selects
Current function if not already selected.
Parameters:
<DNPD>
<UNIT>
·A
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CDC 0.01
CDC? Response: 1.000000E-02 A
[:SOURce]:CDC:RANGe
Syntax:
[:SOURce]:CDC:RANGe <CPD>
[:SOURce]:CDC:RANGe?
Description:
This command selects range in Current function. Node SOUR:CDC also selects Current
function if not already selected.
Parameters:
<CPD>
·AUTO
·50MA
·25MA
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
CDC:RANG AUTO
CDC:RANG? Response: AUTO
[:SOURce]:CDC:LIMiting[:AMPLitude]
Syntax:
[:SOURce]:CDC:LIMiting[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:CDC:LIMiting[:AMPLitude]?
Owner's Manual
output current limiting is always maximum
Range -0.05 ... 0.05 (default 0.01)
{A}
Amper
{AUTO|50MA|25MA} (default AUTO)
auto-range mode
fixed 50 mA range
fixed 25 mA range
OCM160 Precision DC Calibrator
54