OCM160 Precision DC Calibrator
[:SOURce]:FREQuency:DUTY
Syntax:
[:SOURce]:FREQuency:DUTY <DNPD>
[:SOURce]:FREQuency:DUTY?
Description:
This command sets duty cycle of generated signal in Frequency function. Node
SOUR:FREQ also selects Frequency function if not already selected.
Parameters:
<DNPD>
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
FREQ:DUTY 0.5
FREQ:DUTY? Response: 5.000000E-01
[:SOURce]:FREQuency:PULSes[:STATe]
Syntax:
[:SOURce]:FREQuency:PULSes[:STATe] <BOOL>
[:SOURce]:FREQuency:PULSes[:STATe]?
Description:
This commnad switch on pulses generator. Node SOUR:FREQ also selects Frequency
function if not already selected.
Parameters:
<BOOL>
·ON
·OFF
·1
·0
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
FREQ:PULS ON
FREQ:PULS? Response: 1
[:SOURce]:FREQuency:PULSes:COUNt
Syntax:
[:SOURce]:FREQuency:PULSes:COUNt <DNPD>
[:SOURce]:FREQuency:PULSes:COUNt?
Description:
This command sets number of generated pulses. Pulses generator must be set to On (see
FREQ:PULS). Node SOUR:FREQ also selects Frequency function if not already
selected.
Parameters:
<DNPD>
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
61
Range 0.005 ... 0.995 (default 0.5)
{ON|OFF|1|0} (default 0)
pulses generator activate (see FREQ:PULS:COUN)
pulses generator deactivate
same as ON
same as OFF
Range 1 ... 9999999 (default 100)
Orbit Controls AG
Owner's Manual