Orbit Controls AG
7.6.
SCPI Command Tree
This chapter sumarizes all public SCPI commands supported by the device in alphabetic
order. Detailed description follows in next chapter.
:CALibration
:DATE(?) <DNPD>,<DNPD>,<DNPD>
:INTerval(?) <DNPD>
:POINt
:AMPLitude(?) <DNPD>
:CONVerter? <DNPD>
:DATE(?) <DNPD>,<DNPD>,<DNPD>
:MEASure? <DNPD>
:MODE(?) <DNPD>
:SELect(?) <DNPD>
:SAVE
:SECure
:EXIT
:PASSword(?) <DNPD>
:DISPlay
:ANNotation
:CLOCk
:DATE
[:STATe](?) {ON|OFF|1|0}
:TOOLtip
[:STATe](?) {ON|OFF|1|0}
:METer
[:STATe](?) {ON|OFF|1|0}
:BRIGhtness(?) <DNPD>
:LANGuage(?) {ENGLish}
:MEMory
:COUNt? <DNPD>
:PART<IND_MEMORY>
:NAME? <CPD>
:SIZE? <DNPD>
:TRANsfer<IND_COUNTER>(?) <CPD>
:OUTPut
[:STATe](?) {ON|OFF|1|0}
:LOW(?) {FLOat|GROund}
:COMPensation(?) {ON|OFF|1|0}
:RESistance
:SHORt(?) {ON|OFF|1|0}
:COMPensation(?) {ON|OFF|1|0}
:FREQuency
:PULL(?) {ON|OFF|1|0}
[:SOURce]
:VDC
[:AMPLitude](?) <DNPD>[V]
:RANGe(?) {AUTO|100V|30V|3V|300MV}
:LIMiting
[:AMPLitude](?) <DNPD>[A]
:RESet(?) {ADJustable|MAXimum}
Owner's Manual
:FORMat(?) {MDYS|MDYA|DMYS|DMYO|DMYA|YMDS|YMDO}
OCM160 Precision DC Calibrator
36