OCM160 Precision DC Calibrator
Parameters:
<DNPD>
Remarks:
This command requires "Calibration" access
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CAL:POIN:MODE 5
CAL:POIN:MODE? Response: 5
:CALibration:POINt:SELect
Syntax:
:CALibration:POINt:SELect <DNPD>
:CALibration:POINt:SELect?
Description:
This command selects calibration points of particular mode.
Parameters:
<DNPD>
Remarks:
This command requires "Calibration" access
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CAL:POIN:SEL 1
CAL:POIN:SEL? Response: 1
:CALibration:POINt:SAVE
Syntax:
:CALibration:POINt:SAVE
Description:
This command saves current calibration point to non-volatile memory. Unsaved changes
will disappear on restart, selection of another calibration point or close the calibration.
Parameters:
None
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CAL:POIN:SAVE
:CALibration:SECure:EXIT
Syntax:
:CALibration:SECure:EXIT
Description:
This command exits calibration mode and access.
Parameters:
None
Remarks:
The value is set to default after power on
The value is set to default on *RST command
47
Ranges and default values varies in accordance to selected point (see
"Calibration points").
Ranges and default values varies in accordance to selected point (see
"Calibration points").