OCM160 Precision DC Calibrator
Example:
OUTP:LOW FLO
OUTP:LOW? Response: FLO
:OUTPut:COMPensation
Syntax:
:OUTPut:COMPensation <BOOL>
:OUTPut:COMPensation?
Description:
This command select 4-wire voltage outputs connection (remote sense).
Parameters:
<BOOL>
·ON
·OFF
·1
·0
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
OUTP:COMP ON
OUTP:COMP? Response: 1
:OUTPut:RESistance:SHORt
Syntax:
:OUTPut:RESistance:SHORt <BOOL>
:OUTPut:RESistance:SHORt?
Description:
This command turns on short function. "Short" is activated only if output terminals are
switched on (see OUTP:STAT command).
Parameters:
<BOOL>
·ON
·OFF
·1
·0
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
OUTP:RES:SHOR ON
OUTP:RES:SHOR? Response: 1
:OUTPut:RESistance:COMPensation
Syntax:
:OUTPut:RESistance:COMPensation <BOOL>
:OUTPut:RESistance:COMPensation?
Description:
This command select 4-wire resistance outputs connection (remote sense).
Parameters:
<BOOL>
51
{ON|OFF|1|0} (default 0)
selsects 4-wire connection
selsects 2-wire connection
same as ON
same as OFF
{ON|OFF|1|0} (default 0)
short is set if output is on
resistance is set if output is on
same as ON
same as OFF
{ON|OFF|1|0} (default 1)