orbit controls OCM160 Betriebsanleitung Seite 46

Inhaltsverzeichnis

Werbung

Orbit Controls AG
Example:
CAL:POIN:AMPL 1.944
CAL:POIN:AMPL? Response: 1.944000E+00
:CALibration:POINt:CONVerter?
Syntax:
:CALibration:POINt:CONVerter?
Description:
This command reads converter value of particular calibration point.
Parameters:
<DNPD>
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CAL:POIN:CONV? Response: 1.944000E+00
:CALibration:POINt:DATE?
Syntax:
:CALibration:POINt:DATE?
Description:
This command reads date of last calibration of calibration point.
Parameters:
<DNPD>
<DNPD>
<DNPD>
Remarks:
Overlapped command
Example:
CAL:POIN:DATE? Response: 2017,01,13
:CALibration:POINt:MEASure?
Syntax:
:CALibration:POINt:MEASure?
Description:
This command reads measured value of particular calibration point.
Parameters:
<DNPD>
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
CAL:POIN:MEAS? Response: 1.000000E+00
:CALibration:POINt:MODE
Syntax:
:CALibration:POINt:MODE <DNPD>
:CALibration:POINt:MODE?
Description:
This command enters calibration mode. Mode refers to a particular range.
Owner's Manual
Calibration converter value. Ranges and default values varies in
accordance to selected point (see "Calibration points").
Year, Range 2000 ... 2063
Month, Range 1 ... 12
Day, Range 1 ... 31
Calibration measured value. Ranges and default values varies in
accordance to selected point (see "Calibration points").
OCM160 Precision DC Calibrator
46