OCM160 Precision DC Calibrator
7.9.
SCPI Commands
This chapters describes all public SCPI commands in detailed form. The commands here are in
alphabetic order.
:CALibration:DATE
Syntax:
:CALibration:DATE <DNPD>,<DNPD>,<DNPD>
:CALibration:DATE?
Description:
This command sets date of last calibration.
Parameters:
<DNPD>
<DNPD>
<DNPD>
Remarks:
This command requires "Calibration" access
Overlapped command
Example:
CAL:DATE 2017,01,13
CAL:DATE? Response: 2017,01,13
:CALibration:INTerval
Syntax:
:CALibration:INTerval <DNPD>
:CALibration:INTerval?
Description:
This command sets recommended calibration interval for this unit.
Parameters:
<DNPD>
Remarks:
This command requires "Calibration" access
Overlapped command
Example:
CAL:INT 12
CAL:INT? Response: 12
:CALibration:POINt:AMPLitude
Syntax:
:CALibration:POINt:AMPLitude <DNPD>
:CALibration:POINt:AMPLitude?
Description:
This command sets calibration value of particular point. Data will be saved to nonvolatile
memory on CAL:POIN:SAVE command.
Parameters:
<DNPD>
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
45
Year, Range 2000 ... 2063
Month, Range 1 ... 12
Day, Range 1 ... 31
Month, Range 1 ... 240
Calibration point value. Ranges and default values varies in
accordance to selected point (see "Calibration points").