OCM160 Precision DC Calibrator
Description:
This command places device in the REMOTE mode and locks all keys including LOCAL
key. The Command is valid only for RS232, LAN, USB interfaces. The device will not
respond to any other command until is in REMOTE mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:RWL
:SYSTem:TIME
Syntax:
:SYSTem:TIME <DNPD>,<DNPD>,<DNPD>
:SYSTem:TIME?
Description:
This commands allows set system device time (RTC).
Parameters:
<DNPD>
<DNPD>
<DNPD>
Remarks:
Overlapped command
Example:
SYST:TIME 10,45,15
SYST:TIME? Response: 10,45,15
:SYSTem:VERSion?
Syntax:
:SYSTem:VERSion?
Description:
This query retreives version of implemented SCPI language
Parameters:
<CPD>
Remarks:
Overlapped command
Example:
SYST:VERS? Response: 1999.0
:UNIT:TEMPerature
Syntax:
:UNIT:TEMPerature <CPD>
:UNIT:TEMPerature?
Description:
This function allows to set unit for all temperature functions (Platinum, Nickel).
Parameters:
<CPD>
·CEL
·FAR
·K
Remarks:
Example:
77
Hours, Range 0 ... 23
Minutes, Range 0 ... 59
Seconds, Range 0 ... 59
SCPI language version
{CEL|FAR|K} (default CEL)
degrees of Celsius
degrees of Fahrenheit
Kelvin
Overlapped command
Value is not affected by reset
UNIT:TEMP CEL
UNIT:TEMP? Response: CEL