Herunterladen Inhalt Inhalt Diese Seite drucken

orbit controls OCM160 Betriebsanleitung Seite 58

Inhaltsverzeichnis

Werbung

Orbit Controls AG
[:SOURce]:TCPL:TYPE
Syntax:
[:SOURce]:TCPL:TYPE <CPD>
[:SOURce]:TCPL:TYPE?
Description:
This command selects thermocouple type. Node SOUR:TCPL also selects TC function if
not already selected.
Parameters:
<CPD>
·R
·S
·B
·J
·T
·E
·K
·N
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
TCPL:TYPE R
TCPL:TYPE? Response: R
[:SOURce]:TCPL:RJMode
Syntax:
[:SOURce]:TCPL:RJMode <CPD>
[:SOURce]:TCPL:RJMode?
Description:
This command selects manual or external reference junction compensation. Node
SOUR:TCPL also selects TC function if not already selected.
Parameters:
<CPD>
·MANual
·EXTern
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
TCPL:RJM MAN
TCPL:RJM? Response: MAN
[:SOURce]:TCPL:RJAMplitude
Syntax:
[:SOURce]:TCPL:RJAMplitude <DNPD>
[:SOURce]:TCPL:RJAMplitude?
Description:
This command sets/reads reference junction temperature. Node SOUR:TCPL also selects
TC function if not already selected.
Parameters:
<DNPD>
Owner's Manual
{R|S|B|J|T|E|K|N} (default R)
{MANual|EXTern} (default MAN)
manual reference junction is set (see TCPL:RJAM)
external reference junction is set
Range -50.0 ... 150.0 (default 23.0)
OCM160 Precision DC Calibrator
58

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis