Orbit Controls AG
Overlapped command
Example:
FREQ:PULS:COUN 100
FREQ:PULS:COUN? Response: 100
[:SOURce]:FREQuency:PULSes:ACTual?
Syntax:
[:SOURce]:FREQuency:PULSes:ACTual?
Description:
This command reads actual number of generated pulses.
Parameters:
<DNPD>
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
FREQ:PULS:ACT? Response: 0
[:SOURce]:RESistance[:AMPLitude]
Syntax:
[:SOURce]:RESistance[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:RESistance[:AMPLitude]?
Description:
This command sets resistance amplitude in Resistance function. Node SOUR:RES also
selects Resistance function if not already selected.
Parameters:
<DNPD>
<UNIT>
·OHM
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
RES 1000.0
RES? Response: 1.000000E+03 OHM
[:SOURce]:STEP[:STATe]
Syntax:
[:SOURce]:STEP[:STATe] <BOOL>
[:SOURce]:STEP[:STATe]?
Description:
This command turns on Step mode for the appropriate function. Step mode defines of 32
user programmable timing sequencies for each function. Each sequence contains up to
100 steps (amplitude/duration).
Parameters:
<BOOL>
·ON
·OFF
·1
·0
Owner's Manual
Actual number of generated pulses (range 1 ... 9999999)
Range 10 ... 3e5 (default 1000.0)
{OHM}
{ON|OFF|1|0} (default 0)
step mode activate
step mode deactivate
same as ON
same as OFF