Herunterladen Inhalt Inhalt Diese Seite drucken

orbit controls OCM160 Betriebsanleitung Seite 65

Inhaltsverzeichnis

Werbung

OCM160 Precision DC Calibrator
[:SOURce]:STEP:PRESet:ROW<IND_ROW>:AMPLitude?
Description:
This command sets / retrieves selected row in timing table. Data will be saved to
nonvolatile memory on STEP:PRES:SAVE command.
Parameters:
<IND_ROW>
<SPD>
Remarks:
Overlapped command
Example:
STEP:PRES:ROW1:AMPL "0.5,220.0"
STEP:PRES:ROW1:AMPL? Response: "5.000000E-01,2.200000E+02"
[:SOURce]:STEP:PRESet:ROW<IND_ROW>:RDELete
Syntax:
[:SOURce]:STEP:PRESet:ROW<IND_ROW>:RDELete
Description:
This command deletes row from timing table. Data will be saved to non-volatile memory
on STEP:PRES:SAVE command.
Parameters:
<IND_ROW>
Remarks:
Overlapped command
Example:
STEP:PRES:ROW1:RDEL
[:SOURce]:STEP:PRESet:SAVE
Syntax:
[:SOURce]:STEP:PRESet:SAVE
Description:
This command saves current sequence to non-volatile memory. Unsaved changes will
disappear on restart, function change or selection of another sequence.
Parameters:
None
Remarks:
Overlapped command
Example:
STEP:PRES:SAVE
:MEASure:VOLTage[:AMPLitude]?
Syntax:
:MEASure:VOLTage[:AMPLitude]?
Description:
This query returns value of output voltage in Current function.
Parameters:
<DNPD>
<UNIT>
·V
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
65
Range 1 ... Row count (1 - if omitted)
Quoted string representing amplitude. The amplitude consists of two
float numeric fields separated by comma. The first one represents
timing interval in seconds and the second one amplitude.
Range 1 ... Row count (1 - if omitted)
Range -100.0 ... 100.0
{V}
Voltage
Orbit Controls AG
Owner's Manual

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis