OCM160 Precision DC Calibrator
:MEASure:COUNter[:AMPLitude]?
Syntax:
:MEASure:COUNter[:AMPLitude]?
Description:
This query returns value of counter.
Parameters:
<DNPD>
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
MEAS:COUN? Response: 1.0E+006
:MEASure:TCPL:RJAMplitude
Syntax:
:MEASure:TCPL:RJAMplitude?
Description:
This query returns external reference junction temperature.
Parameters:
<DNPD>
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
MEAS:TCPL:RJAM? Response: 2.300000E+01
:STATus:OPERation:CONDition
Syntax:
:STATus:OPERation:CONDition?
Description:
This query returns the content of Operational Condition register. It is a decimal value
which corresponds to the binary-weighted sum of all bits in the register. Register is not
cleared after this query. The response to the query therefore represents an instantaneous
'Snapshot' of the register state, at the time that the query was accepted.
Parameters:
<DNPD>
Remarks:
Overlapped command
Example:
STAT:OPER:COND? Response: 2
:STATus:OPERation:ENABle
Syntax:
:STATus:OPERation:ENABle <DNPD>
:STATus:OPERation:ENABle?
Description:
This command enables bits in the Operational Data Enable register. Selected bits are
summarized at bit 7 (OSS) of the IEEE488.2 Status Byte register.
Parameters:
<DNPD>
Remarks:
Overlapped command
67
Measured value
Measured value
Operational Condition register
Operational Data Enable register
Orbit Controls AG
Owner's Manual

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis