1. Enter "Measuring Point (MP)" menu. See Chapter 4 on Page 50.
⇒ The screen "MP Select" with the item "New BT device recognized 00" is shown.
2. Tap "+" for the item "New BT device recognized 00".
3. Configure the new measuring point (MP).
4. Tap "OK"
⇒ The screen "MP Select" is shown.
5. Activate the radio button.
6. Return to the next-higher menu level.
7. Answer the question "Save changes?" with "Yes".
⇒ The Measurement Screen for the selected measuring point is shown. After some seconds
the measurement results are shown. Tap "Save" to save the results on the USB stick. Tap
"Repeat" to get new values from the sensor. Tap "Cancel" to cancel the procedure.
7

Calibration

You can also calibrate the dismounted InTap sensor or the dismounted In-Line Sensor via
iSense™ software. See Operating Instructions of the iSense software.
As soon as the calibration is in progress no other calibration can be started.
Calibration Menu c
After every successful calibration different options are available. If "Adjust" or "Calibrate" is
chosen, the message "Calibration saved successfully!" is displayed. Press "Done".
Option
Adjust
Calibrate
Cancel
en
54
ISM sensors
Calibration values are stored in the sensor and used for the measurement.
Additionally, the calibration values are stored in the calibration history.
Calibration values are stored in the calibration history for documentation, but not
be used for the measurement. The calibration values from the last valid
adjustment are further used for the measurement.
Calibration values are discarded.
InTap Portable oDO Analyzer • 30 461 735 A