Calibration

6.
Calibration
6.1

Settings

When using the SevenEasy conductivity meter, you only need to do a 1-point calibration or enter a cell
constant directly.
Press the
key. The current calibration setting starts to blink.
Use the
or
key to select other items either above or below. When the desired option blinks,
press
to confirm your selection.
If you choose to set the cell constant, the current setting appears and the first digit blinks. Use the
and
keys to increase or decrease the value. Press the
your setting, follow the same procedure to set the next digits. Press
default setting is 1.000.
6.2
Calibration
Place the conductivity sensor in a calibration standard and press
The SevenEasy conductivity meter automatically endpoints when calibrating. To manually endpoint,
press
. The meter displays and freezes the standard value.
To return to sample measurement, press
Note
- If you have entered the cell constant of your sensor and choose to use it for your measurement,
you actually do not need to perform a calibration with a standard. If you press the
this circumstance, the instrument shows the currently entered cell constant's value instead of
performing a calibration.
- To ensure the most accurate conductivity readings, you need to perform a calibration regularly.