Problem
Inefficient therapy
Table 4 – Quick Troubleshooting
Possible Cause
Insufficient pressure in the
air supply circuit
Insufficient or excessive pow-
der level in the container
Polisher nozzle clogged due
to an excessive amount of
moisture in the powder or
insufficient
cleaning/maintenance
Irregular and/or intermittent
water/powder flow
The starjet device is tilted
more than 45° with respect
to the correct position
TROUBLESHOOTING
Solution
Check the pressure of the
dental unit air circuit
Restore the correct level of
powder in the container (see
Chapter 5.2 on page 14
point 9)
Read
Chapter 5.3 on page
17
,
Chapter 6.1 on page
21
,
Chapter 6.5 on page
25
and
Chapter 7 on page
37
.
Remove the powder from the
container and clean it using
a dry cloth
Check the O-ring of the
polisher nozzle coupling (see
Chapter 7.1 on page 37
Correctly reposition the
grip of starjet and use the
dedicated polisher nozzles
for 360° rotation without
rotating the entire device
,
)
43