Herunterladen Inhalt Inhalt Diese Seite drucken

Daily Decontamination Of The Dental Unit Water Circuit; Cleaning The Sterilisable Accessories - mectron starjet Gebrauchs- Und Wartungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 116
CLEANING, DISINFECTION AND STERILISATION
6.4
Daily Decontamination of the Dental Unit Water
Circuit
This procedure must be carried out:
• at the start of the day;
• after each patient;
• at the end of the day, before stowing the
device.
To prevent the spread of bacteria in the wa-
ter circuit of the dental unit to which starjet is
connected, fill the dental unit water circuit
tank with a chlorine dioxide solution at 0.1%
to treat the dental unit lines, diluted with tap
water at a ratio of at least 1:10 (as per the
manufacturer's instructions
h) Procedure validated using the product MicroCLEAR®
Dental Unit Waterline Cleaner.
6.5
Cleaning the Sterilisable Accessories
The sterilisable parts of the appliance are:
• Polisher Nozzles;
• K9 Wrench;
• Cleaning Needles.
Before proceeding with operations to check
cleanliness (
Chapter 6.6 on page 34
(
Chapter 6.7 on page 35
sation (
Chapter 6.8 on page 35
three possible cleaning methods explained in
detail in the following sub-chapters must be
selected.
solution, carefully following the
).
d)
manufacturer's instructions.
It is advisable to empty and fill the dental unit
tank with a fresh solution every day.
have been validated by MECTRON in order
to enable the medical device to be reused.
The person responsible for the maintenance,
cleaning and sterilisation process, must
ensure the that repeated processes are
effectively carried out using the equipment,
), drying
materials and staff in the reprocessing
) and then sterili-
structure, in order to achieve the desired
), one of the
result. This generally requires the validation
and systematic monitoring of the process.
Similarly, all deviations from the instructions
provided by the person responsible for the
processes should be properly evaluated for
judging the effectiveness and the potential
undesirable consequences.
1.
Prepare the decontamination solution
by adding the required volume of
concentrate to the dental unit water
tank (volume required as per the
manufacturer's instructions).
2.
Fill the rest of the tank with tap water;
3.
Operate the pedal and run the solution
through the device for at least two
minutes.
WARNING: Prepare the decontamination
CAUTION: The following instructions
25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis