Remove the K9 wrench from the enzy-
matic detergent solution.
Thoroughly rinse and brush all the sur-
faces of the K9 wrench (see previous
point) under running water for at least 10
minutes;
Place the K9 wrench in the ultrasonic tank
immersed in the enzymatic detergent
solution at 24 °C ±2 °C and run a cycle
for at least 20 minutes;
Remove the K9 wrench from the ultrason-
ic tank and rinse it under running water
so as to eliminate all detergent residue.
Brush the internal and external surfaces
of the K9 wrench with a clean brush with
soft nylon bristles, under running water.
6.5.1.4
Cleaning Needles
Place the cleaning needle in a clean con-
tainer, in a horizontal position, and add
enough enzymatic detergent solution to
completely cover the cleaning needle;