starjet
6.5.1
Manual Cleaning
Manual cleaning can be carried out as an alternative to the automatic cleaning described in
Chapter 6.5.2 on page 33.
6.5.1.1
Required Materials
• Enzymatic detergent at pH 6-9;
• Water;
• Container for immersion in the
enzymatic liquid;
• Ultrasonic tank;
Obtain a bucket and ultrasonic tank large
enough to hold the accessories needing
to be cleaned.
Prepare an enzymatic detergent solutio-
n
at pH 6-9 according to the manufac-
a)
turer's instructions, in a sufficient quan-
tity to fill the bucket and ultrasonic tank.
Use warm water for the bucket (30°C -
35°C) and water at ambient temperature
(22°C ±2°C) for the ultrasonic tank.
CAUTION: Once used, dispose of the
enzymatic detergent correctly, do not
recycle it.
a) Process validated by independent bodies with
Mectron ENZYMEC enzymatic detergent, 0.8%
v/v.
6.5.1.2
Polisher Nozzles
Free the channel of the polisher nozzle of
any powder residue using the supplied Ø
0.4 mm cleaning needle, from both sides.
Blow compressed air into the central hole
in the tip, from both ends;
CAUTION: Clean the channel of the
nozzle using only the Ø 0.4 mm cleaning
needle supplied with the appliance.
CAUTION: If the SUBGINGIVAL
PERIO TIP is on the Polisher nozzle,
remove and dispose of it.
26
• Clean, soft, low-lint cloths;
• Brush with soft nylon bristles;
• 25 ml syringe;
• Demineralised water.
1
2