Risk of injury caused by insufficiently tightened fixing screws of the polyaxial
elements!
Using a torque wrench, tighten the fixing screws to a torque of 10 Nm.
The screwdriver for the retaining bolt must engage fully in the fixing screw to
WARNING
prevent it from slipping.
The groove on the screwdriver to accommodate the retaining bolt must be
completely covered by the guide tube.
Risk of injury if the centralizer is left in the body!
In the event of failure of the 10 Nm torque wrench, the retaining bolt can be
locked using a ¼" T-handle grip.
WARNING
Operating time may be increased as a result of incorrect application!
Do not pull out the insertion instrument after the first click!
Only pull out the insertion instrument after completion of the second click.
Ensure that the insertion instrument does not slip off the tensioning element after
WARNING
the first click.
Further information about implant systems can be obtained at all times from B. Braun/Aesculap or the relevant
B. Braun/Aesculap subsidiary.