10. SETUP-Additional menu items
GB
10.1 SETUP-Additional menu items: Functions
a -->
b -->
c --{
d -->
e -->
a) Ref:
Current measured value of the connected CPT6000 pressure reference sensor
b) Tare:
Offset-Function, this adjusts the current measured value. The value entered is
added to the current measured value.
(e.g. Ref. 0.000 & Tara: 1.000 --> [new] Ref. 1.000)
c) Min:/Max:
Minimum and Maximum-value memory
The memory is reset by highlighting the value with the cursor
pressing the Clear button.
d) Alarm:
Audible and visible Alarm function.
upper alarm limit: > = x bar
lower alarm limit: < = x bar
If the current measured value goes over the set alarm limit, an intermittent alarm
tone sounds and the lower display line blinks.
Activated via:
Move the cursor to the field next to the word alarm that reads "<OFF>" and
via
toggle it to "<ON>".
Deactivated via:
Toggle back to "<OFF>"
44
Choice from menu item
Input via numeric keypad
Confirmation of the input
ENTER
(Clear delet the input and the
Min/Max memory is reset)