8. Display description

GB

8.1 Device status message briefly after switching on the CPH6000

Immediately after switching on the instrument, the following status messages
appear briefly:
a) The 24 V supply voltage (accessed from the top edge of the instrument) can be
switched on or off during the configuration of each operating mode. If it is not
needed for a measurement, then it should be switched off, in order to save energy.
b) Current battery level (see Chapter 10.5)
c) Height difference in [mm]
In the "SETUP \ CPH-Settings" menu the height difference between the test
specimen and the CPT6000 pressure sensor can be set. This value prompts
an automatic correction calculation based on a head pressure, which
compensates for the pressure difference. This value must be correct for the
following measuring procedure and/or be altered accordingly in the
"SETUP \ CPH-Settings" menu (see chapter 5.2).
d) Temperature in [°C]
In the "SETUP \ CPH-Settings" menu the temperature (ambient temperature)
can be entered, which is also later recorded on the calibration certificate. This
value must be correct for the following measuring procedure and/or be altered
accordingly in the "SETUP \ CPH-Settings" menu (see chapter 5.2)
e) Current date of the integrated real-time clock
The date in the real-time clock, which is also later noted within the calibration
certificate, is adjusted via the "SETUP \ CPH-Settings" menu. This value must
24