h) Offset:
This menu option only appears if the CPH6000s reference pressure sensor is an
absolute pressure sensor. Through this menu option the measured value of the
reference pressure sensor can be adjusted. This should only be used, however,
as close as possible to the absolute zero, and using a reference that is at least 4
times more accurate than the CPH6000s reference sensor.
i) Calibrated:
Calibration Date of the CPT6000 sensor (Year/Month/Day)
10.4 SETUP-Additional menu items: Ref. Sensor list
a -->
b ->
c ->
The instrument supports up to 5 CPT6000 Sensors and these are listed in this
menu.
a) Currently connected CPT6000 reference pressure sensor
b) Sensor-List of the supported CPT6000-Sensors (calibrated in conjunction with
the instrument)
c) Data of the particular sensor, which has been selected using the cursor
(Date: Year/Month/Day)
WIKA Operating Instructions ProcessCalibrator
Choice
one of the listed sensors
with
GB
47