ProcessCalibrator
CPH6000

5.3 Important device settings for calibrations using calibration mode

GB
Calibration date
The instrument has an integral real-time clock with date. The current date of a
calibration is stated later in the calibration certificate. Before starting a
calibration you must ensure that the internal date of the CPH6000 is correct
(see Chapter 10.5 "SETUP-Additional menu options: CPH-Settings").
Note: After switching on the CPH6000, a brief device status message
appears showing the current settings.
Calibration temperature
On calibration certificates for pressure measuring instruments, the ambient
temperature during the calibration is also always stated along with the calibra-
tion data. Therefore, the current ambient temperature can be entered into the
CPH6000 via its menu so that it can later be included on the calibration certifi-
cate. Before starting a calibration you must ensure that the temperature value
set in the CPH6000 corresponds to the ambient temperature (see Chapter
10.5 "SETUP-Additional menu options: CPH-Settings").
Note: After switching on the CPH6000, a brief device status message
appears showing the current settings.

5.4 Unit and resolution

After selecting one of the Main Menu Items (e.g.: MEASURING, CALIBRATION
or SWITCH TEST) from the "SETUP" menu (press "SETUP" key), using the menu
Option: "Unit", and its associated submenu respectively (move the cursor to "Unit"
and press the right or left arrow), you can select the unit and adjust its resolution
(see Chapter 9: "Operating modes").