ProcessCalibrator
CPH6000
9.2 MEASURING mode (with test item)
To switch the instrument into "MEASURING" mode (with test item display the test
signal as an electrical signal or as a pressure), in order to carry out a comparati-
ve measurement / Calibration without measured-value recording, then follow the
instructions on the previous page.
The following is a more detailed explanation of Point: "2. Preparation for
MEASUREMENT"
a: Test item and respective test item signal [0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V /
0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or mechanical for pressure gauges]
If a comparative measurement with a mechanical gauge (test item) is being
made, then the gauge's measured value should be entered via the numeric
keypad and confirmed with ENTER.
b: Lower limit of test item measuring range
c: Upper limit of test item measuring range
d: Measurement accuracy of the test item in % FS (with respect to the span) or
% rd (with respect to the measured value)
e: Measurement accuracy of the test item in %
f: Units & resolution (Sub-menu)
g: Measurement type for the test item (relative or absolute)
h: Test medium (pneumatic -> gas & hydraulic -> oil)
i: Power supply for test item (on/off) [If no external supply is required for the
test item, "OFF" should be selected to conserve energy]
xxx
Current coursor position; alter via
} Parameter-selection from list or menu via
0.00
Parameter input using numeric keypad
Input confirmation
ENTER
WIKA Operating Instructions ProcessCalibrator
Choose & select (standard units)
via
SELECT
Customer-specific units; with respect to
bar (Input via numeric keypad)
Display resolution in operating mode via
(back with
BACK
Clear input
CLEAR
)
33
GB