WIKA CPH6000 Betriebsanleitung Seite 41

Prozesskalibrator
Vorschau ausblenden Andere Handbücher für CPH6000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ProcessCalibrator
CPH6000
9.3 CALIBRATION mode (Calibration of a pressure gauge)
1st Test point
(e.g. 0 bar)
Reference ->
Test item ->
Deviation ->
Test item no. & ->
Test point no.
2nd Test point
(e.g. 1 bar)
Reference ->
Test item ->
Deviation ->
Test item no. & ->
Test point no.
2nd Test point
Reference ->
Test item ->
Deviation ->
Test item no. & ->
Test point no.
xth Test point
WIKA Operating Instructions ProcessCalibrator
Generate the specified Set-
point acc. to the test item
using the pressure generator
If the set value = 0 bar esta-
blish a pressure free condi-
tion/Atmosphere (test item
must indicate 0 bar; zero-point
correction if necessary) and
with
<---
ENTER
record the values of this test
point
Generate the specified Set-
point acc. to the test item
using the pressure generator
record the values of this test
ENTER
point
(using
returns you to the
BACK
previous test point)
GB
41