6. Commissioning, operation

6.2 Explanation of the display

6.2.1 Instrument status messages shortly after switching on the CPH7650
Directly after the instrument is switched on, the following status messages are displayed
briefly:
a) The voltage supply of 24 V (available at the upper end of instrument) can, during the
configuration of each operating mode, be switched on or off. If it is not needed for a
measurement, then it should be switched off, in order to conserve energy.
b) Current battery capacity (see chapter 6.4.5 "SETUP additional menu items: CPH
configuration")
c) Height difference in mm
In the "SETUP\CPH configuration" menu, set the height difference between the test
item and the CPT6000 reference pressure sensor. This value influences an automatic
correction calculation in order to eliminate any pressure difference based on a
medium column. This value must be correct for the following measuring procedure
and/or be adjusted accordingly in the "SETUP/CPH configuration" menu (see chapter
6.4.5 "SETUP additional menu items: CPH configuration").
d) Temperature in [°C]
In the "SETUP\CPH configuration" menu, the temperature (ambient temperature) is
entered. This value can be adjusted accordingly in the "SETUP\CPH configuration"
menu (see chapter 6.4.5 "SETUP additional menu items: CPH configuration").
e) Current date of the integrated real-time clock
In the "SETUP\CPH configuration" menu, the date of the real-time clock is set, which is
later marked on the calibration certificate. This value must be correct for the following
measuring procedure in calibration mode and/or be adjusted accordingly in the
"SETUP/CPH configuration" menu (see chapter 6.4.5 "SETUP additional menu items:
CPH configuration").
f) Calibration date for the electrical measuring inputs of the CPH7650 (Year/Month/Day)
Following the status messages, the display returns to the screen for the last-selected
operating mode (see following chapter 6.2.3 "Display contents of the operating
modes").
WIKA operating instructions, model CPH7650
CPH7650
a
b
c
d
e
f
EN
25