8.2.2 Operating mode: CALIBRATION
In the CALIBRATION mode the data shown above the (dashed dividing) line is the
same as in the "MEASURING with test item" Mode (see 8.2.1).
CPT6000 reference pressure sensor
Deviation between reference and test item
ID-No: test item IDENT number
8.2.3 Operating mode: SWITCH TEST
The SWITCH TEST mode, along with the reference pressure sensor data (see
MEASURING mode), displays the Status and Switching Points of the pressure
switch.
CPT6000 reference pressure sensor
Current switch position / pressure
Hysteresis / distance between
Switch opening and closing point
WIKA Operating Instructions ProcessCalibrator
Test item
Calibration set-point
True calibration value
P-01: test item No. 1
<01>: test point No. 1
True test item signal
switch status
Opening switching point
Closing switching point
{
{
}
Pressure signal (test item)
Electr. signal: test item
{
GB
P
switchable
I/U
pressure unit
(can be set
via menu)