6. Commissioning, operation

6.7 SETUP additional menu items

6.7.1 SETUP additional menu items: Functions

..Functions
1
Ref.:
2
Tare:
Min:
3
Max:
4
Alarm:
5
Filter:
Ref. = rel. Pressure
Offset pressure
Current measured value of the connected CPT6000 reference pressure sensor
1
Offset function that influences the current measured value. The value entered is
2
added to the current measured value.
(e.g. Ref. 0.000 and Tare: 1.000 → [new] Ref. 1.000)
Minimum and maximum value memory
3
The memory is reset by highlighting the value with the [▲] or [▼] cursor and pressing
the [CLEAR] button.
Audible and visible alarm function
4
upper alarm limit: ≥ bar
lower alarm limit: ≤ bar
If the current measured value goes outside the set alarm limits, an intermittent alarm
tone sounds and the lower status line blinks.
Activation via:
Move the cursor to the field "OFF" next to the word Alarm and via [◀] or [▶] change
it to "ON".
Deactivation via:
setting it back to "OFF"
Filter [1-5]:
5
Damping/smoothing of the reference sensor signal
Definition of the numbers:
1 = no additional smoothing ... 5 = strong smoothing
WIKA operating instructions, model CPH7650
bar
0.000
0.000
bar
0.000
bar
bar
0.000
ON
10.00
bar
bar
-1.00
0
Select menu item
Input via numeric keypad
ENTER
Confirmation of input with
[ENTER]
([CLEAR] deletes the input, or
resets the MIN/MAX memory)