9.3 CALIBRATION mode
In order to put the instrument into "CALIBRATION" Mode, the procedure on the
previous page should be followed.
The following is a more detailed explanation of Point: "2. Preparation for CA-
LIBRATION"
a: No. of the calibration, and therefore the test item (up to 16 calibrations, each
with up to 32 test points, can be predefined and then recorded).
b: Test Item and respective Test Item signal [0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V /
0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or mechanical for pressure gauges]
c: ID-No. of the test item
d: Measuring point No. of the test item
e: Lower limit of test item measuring range
f: Upper limit of test item measuring range
g:h: Measurement accuracy of the test item in % FS (with respect to the span) or
% rd (with respect to the measured value)
i: Units & resolution (Sub-menu)
j: Measurement type for the test item (relative or absolute)
k: Test medium (pneumatic -> gas & hydraulic -> oil)
l: Power supply for test item (on/off) [If no external supply is required for the
test item, "OFF" should be selected to conserve energy]
m: No. of test point x
n: Optional "Delay Time" [sec] to conform with DKD requirements. (see under 9.3)
o: Value of test point x (Input via numeric keypad)
(Test point x+1 and x-1 can be accessed respectively using "
p: True value of the test item (will be recorded during the calibration)
xxx
Current coursor position; alter via
} Parameter-selection from list or menu via
0.00
Parameter input using numeric keypad
Input confirmation
ENTER
WIKA Operating Instructions ProcessCalibrator
Choose & select (standard units)
via
SELECT
Customer-specific units; with respect to
bar (Input via numeric keypad)
Display resolution in operating mode via
(back with
)
BACK
Clear input
CLEAR