10.5 SETUP-Additional menu items: CPH-Settings
GB
a -->
b -->
c -->
d -->
e -->
f -->
g -->
h -->
i -->
This Menu section lists general instrument settings, such as:
a) Input for ambient temperature, that is later recorded on the calibration
certificate.
b) Input for height difference between Reference-Pressure sensor and test item
used in the automatic correction (deduction of a head pressure)
c) Selection of the menu language (German/English)
Note: Changing the language takes approx. 3 seconds.
d) System Clock Date (Year/Month/Day)
e) System Clock Time (Hour/Minute/Second)
f) Brightness of the display backlight
g) Display Contrast (not adjustable)
h) Powersave-Function (automatic switch-off delay for backlight and internal 24 V
test item power supply). If the instrument is idle for the set switch-off delay
period (no keys pressed and no digital communication), then the backlight and
the 24 V test item power supply will be switched off, until any key is pressed
(except ON/OFF) or the instrument is addressed over the digital interface.
48
Choice from menu item
Input via numeric keypad
Confirmation of the input
ENTER
(Clear delet the input and the
memory is reset)