10.3 SETUP-Additional menu items: Reference Sensor
GB
a) Sensor No:
Serial Number of the currently connected CPT6000 reference pressure sensor
b) R - Start / Range-End:
Upper and lower limits of the measuring range of the currently connected
CPT6000 sensor
c) Units:
Basic pressure unit of the CPT6000 sensor
d) Class:
Measuring chain accuracy of the CPH6000 with connected CPT6000 sensor
e) Measurement Type:
Pressure type of the currently connected CPT6000 sensor (gauge or absolute
pressure)
f) Overload:
Length of time for which the CPT6000-Sensor was unacceptably overloaded.
If in this case the unpressurised instrument no longer reads zero, then
it is highly probable that the instrument no longer meets its specified
accuracy class. The only solution for this is an immediate
recalibration. (For absolute pressure sensors < 1 bar this function is
deactivated, since for this measuring range atmospheric pressure
already represents an overload)
g) Ref Value:
Current measured value of the connected CPT6000 reference pressure sensor